Temperature change in pig rib bone during implant site preparation by low-speed drilling
DOI:
https://doi.org/10.1590/S1678-77572010000500016Keywords:
Implant drilling, Infrared thermography, Heat production, Bone necrosisAbstract
OBJECTIVES: The purpose of this study was to evaluate the temperature change during low-speed drilling using infrared thermography. MATERIAL AND METHODS: Pig ribs were used to provide cortical bone of a similar quality to human mandible. Heat production by three implant drill systems (two conventional drilling systems and one low-speed drilling system) was evaluated by measuring the bone temperature using infrared thermography. Each system had two different bur sizes. The drill systems used were twist drill (2.0 mm/2.5 mm), which establishes the direction of the implant, and finally a 3.0 mm-pilot drill. Thermal images were recorded using the IRI1001 system (Infrared Integrated Systems Ltd.). Baseline temperature was 31±1ºC. Measurements were repeated 10 times, and a static load of 10 kg was applied while drilling. Data were analyzed using descriptive statistics. Statistical analysis was conducted with two-way ANOVA. RESULTS AND CONCLUSIONS: Mean values (n=10 drill sequences) for maximum recorded temperature (Max TºC), change in temperature (ΔTºC) from baseline were as follows. The changes in temperature (ΔTºC) were 1.57ºC and 2.46ºC for the lowest and the highest values, respectively. Drilling at 50 rpm without irrigation did not produce overheating. There was no significant difference in heat production between the 3 implant drill systems (p>;0.05). No implant drill system produced heat exceeding 47ºC, which is the critical temperature for bone necrosis during low-speed drilling. Low-speed drilling without irrigation could be used during implant site preparation.Downloads
Downloads
Published
2010-10-01
Issue
Section
Original Articles
License
Todo o conteúdo do periódico, exceto onde está identificado, está licenciado sob uma Licença Creative Commons do tipo atribuição CC-BY.
How to Cite
Kim, S.-J., Yoo, J., Kim, Y.-S., & Shin, S.-W. (2010). Temperature change in pig rib bone during implant site preparation by low-speed drilling . Journal of Applied Oral Science, 18(5), 522-527. https://doi.org/10.1590/S1678-77572010000500016