A rationale method for evaluating unscrewing torque values of prosthetic screws in dental implants

Authors

  • Felipe Miguel Saliba Estácio de Sá University; Dental School; Department of Prosthodontics
  • Mayra Cardoso State University of Rio de Janeiro; Dental School; Department of Prosthodontics
  • Marcelo Ferreira Torres State University of Rio de Janeiro; Dental School; Department of Prosthodontics
  • Alexandre Carvalho Teixeira State University of Rio de Janeiro; Dental School; Department of Prosthodontics
  • Eduardo José Veras Lourenço State University of Rio de Janeiro; Dental School; Department of Prosthodontics
  • Daniel de Moraes Telles State University of Rio de Janeiro; Dental School; Department of Prosthodontics

DOI:

https://doi.org/10.1590/S1678-77572011000100013

Keywords:

Dental implants, Biomechanics, Torque

Abstract

OBJECTIVES: Previous studies that evaluated the torque needed for removing dental implant screws have not considered the manner of transfer of the occlusal loads in clinical settings. Instead, the torque used for removal was applied directly to the screw, and most of them omitted the possibility that the hexagon could limit the action of the occlusal load in the loosening of the screws. The present study proposes a method for evaluating the screw removal torque in an anti-rotational device independent way, creating an unscrewing load transfer to the entire assembly, not only to the screw. MATERIAL AND METHODS: Twenty hexagonal abutments without the hexagon in their bases were fixed with a screw to 20 dental implants. They were divided into two groups: Group 1 used titanium screws and Group 2 used titanium screws covered with a solid lubricant. A torque of 32 Ncm was applied to the screw and then a custom-made wrench was used for rotating the abutment counterclockwise, to loosen the screw. A digital torque meter recorded the torque required to loosen the abutment. RESULTS: There was a significant difference between the means of Group 1 (38.62±6.43 Ncm) and Group 2 (48.47±5.04 Ncm), with p=0.001. CONCLUSION: This methodology was effective in comparing unscrewing torque values of the implant-abutment junction even with a limited sample size. It confirmed a previously shown significant difference between two types of screws.

Downloads

Download data is not yet available.

Downloads

Published

2011-02-01

Issue

Section

Original Articles

How to Cite

Saliba, F. M., Cardoso, M., Torres, M. F., Teixeira, A. C., Lourenço, E. J. V., & Telles, D. de M. (2011). A rationale method for evaluating unscrewing torque values of prosthetic screws in dental implants . Journal of Applied Oral Science, 19(1), 63-67. https://doi.org/10.1590/S1678-77572011000100013