Diagnosis of human toxoplasmosis by a dot enzyme-linked immunosorbent assay

Authors

  • Mohamed Mohamed Anwar Elsaid UFMG; ICB; Departamento de Parasitologia
  • Maria Terezinha Bahia Universidade Federal de Ouro Preto
  • George Luiz Machado-Coelho Universidade Federal de Ouro Preto
  • Ricardo Wagner de Almeida Vitor UFMG; ICB; Departamento de Parasitologia

Keywords:

Human toxoplasmosis, Dot-ELISA, Serodiagnosis

Abstract

A Dot enzyme-linked immunosorbent assay (Dot-ELISA) was standardized and evaluated for the serodiagnosis of human toxoplasmosis. Out of 538 serum samples tested by the immunofluorescence test for toxoplasmosis (IFAT-IgG) as reference test, 183 (34%) were positive at cut off 1:16 and 192 (36%) were positive for Dot-ELISA-IgG at cut-off 1:256. For Dot-ELISA, co-positivity was 0.94, co-negativity 0.94 and concordance 0.88 in relation to IFAT-IgG. These results suggest the usefulness of Dot-ELISA (cut-off titer of 1:256) for the serodiagnosis of human toxoplasmosis. The main advantage of this technique is simplicity, positive test can be visually identified (colored precipitate). It does not require a special equipment and it can be used as a qualitative test to screen large numbers of samples or as a quantitative assay to determine end-point titration of individual sera.