Evaluation of cleaning and disinfection performance of automatic washer disinfectors machines in programs presenting different cycle times and temperatures
DOI:
https://doi.org/10.1590/S0104-11692006000500015Keywords:
disinfection, validation studies^i1^spublication t, colony count^i1^smicrob, technology^i1^sinstrumentatAbstract
Thermal washer-disinfectors represent a technology that brought about great advantages such as, establishment of protocols, standard operating procedures, reduction in occupational risk of a biological and environmental nature. The efficacy of the cleaning and disinfection obtained by automatic washer disinfectors machines in running programs with different times and temperatures determined by the different official agencies was validated according to recommendations from ISO Standards 15883-1/1999 and HTM2030 (NHS Estates, 1997) for the determining of the Minimum Lethality and DAL both theoretically and through the use with thermocouples. In order to determine the cleaning efficacy, the Soil Test, Biotrace Pro-tect and the Protein Test Kit were used. The procedure to verify the CFU count of viable microorganisms was performed before and after the thermal disinfection. This article shows that the results are in compliance with the ISO and HTM Standards. The validation steps confirmed the high efficacy level of the Medical Washer-Disinfectors. This protocol enabled the evaluation of the procedure based on evidence supported by scientific research, aiming at the support of the Supply Center multi-professional personnel with information and the possibility of developing further research.Downloads
Download data is not yet available.
Downloads
Published
2006-10-01
Issue
Section
Original Articles
License
RLAE’s authorship concept is based on the substantial contribution by each of the individuals listed as authors, mainly in terms of conceiving and planning the research project, collecting or analyzing and interpreting data, writing and critical review. Indication of authors’ names under the article title is limited to six. If more, authors are listed on the online submission form under Acknowledgements. The possibility of including more than six authors will only be examined on multicenter studies, considering the explanations presented by the authors.Including names of authors whose contribution does not fit into the above criteria cannot be justified. Those names can be included in the Acknowledgements section.
Authors are fully responsible for the concepts disseminated in their manuscripts, which do not necessarily reflect the editors’ and editorial board’s opinion.
How to Cite
Bergo, M. do C. N. C. (2006). Evaluation of cleaning and disinfection performance of automatic washer disinfectors machines in programs presenting different cycle times and temperatures. Revista Latino-Americana De Enfermagem, 14(5), 735-741. https://doi.org/10.1590/S0104-11692006000500015