Standardization of method for determining glycosylated hemoglobin (Hb A1c) by cation exchange high performance liquid chromatography
DOI:
https://doi.org/10.1590/S1984-82502009000400007Keywords:
High performance liquid chromatography, Mono S^i1^smet, Hemoglobin^i1^s, Hemoglobin^i1^svariaAbstract
Hemoblobin A1c is the most important parameter for the monitoring of metabolic control of patients with diabetes mellitus. The purpose of this study was to adapt the Mono S method to a conventional HPLC system, allowing highly selective HbA1c determination without the acquisition of kits or the use of dedicated systems The results obtained were compared to the Tinaquant® immune turbidimetric method and the Bio-Rad Variant® chromatographic method. The developed method presented intra-study precision (C.V. %) of 1.39-3.65 and inter-study precision (C.V. %) of 2.80-3.02%. The determination coefficients among methods were: HPLC Mono S x Tinaquant®: r²: 0.9856 (n=60) and HPLC Mono S x HPLC Bio-Rad Variant®: r²: 0.9806 (n=16). A conversion equation between HPLC Mono S and Bio-Rad Variant® was calculated allowing yielding comparable and interchangeable values. The HPLC Mono-S is a precise, low-cost method which yields similar values to the Bio-Rad Variant® method on conventional HPLC equipment.Downloads
Download data is not yet available.
Downloads
Published
2009-12-01
Issue
Section
Original Papers
License
All content of the journal, except where identified, is licensed under a Creative Commons attribution-type BY.
The on-line journal has open and free access.
How to Cite
Standardization of method for determining glycosylated hemoglobin (Hb A1c) by cation exchange high performance liquid chromatography . (2009). Brazilian Journal of Pharmaceutical Sciences, 45(4), 650-657. https://doi.org/10.1590/S1984-82502009000400007